>> P.11

10-45to+180℃TEMPERATURECHAMBERARU-0680ARU-1100BalancedTemperatureControl(BTC)system-45to+180℃(-49to+356℉)6.3K/min4.8K/minWithin30min.Within50min.±0.3K3.0K+20to+180℃+20to-45℃4.7K/min4.4K/minWithin40min.Within50min.Testareatemperature:+20℃4500W18Cr-stainlesssteelplate(Hairlinenish)18-8Cr-NiStainlesssteelplate(BAnish)NichromestripwireheaterPlatencoolerMechanicalsingle-stagerefrigerationsystemAir-cooledcondenserElectronicexpansionvalveR-404A[R-449A]*3SiroccofanTemp.rangeTemp.uctuationTemp.variationinspaceTemp.rateofchange*2HeatupratePulldownrateTemperatureextremesachievementtimeHeatuptimePulldowntimeAllowableheatloadExteriormaterialTestareamaterialHeaterCoolerSystemCondenserExpansionsystemRefrigerantRefrigerationunitAircirculatorTemp.performance*1ConstructionARUModelSystemCapacityChambertotalloadresistanceInsidedimensionsmm*4Outsidedimensionsmm*4Weight680L80kgW850×H1000×D800W1050×H1955×D1805505kg1100L150kgW1100×H1000×D1000W1300×H1955×D2005595kgAllowableambientconditions0to+40℃(+32to+104℉)/Upto75%rh53A49A23A22A61dBPowersupply*5200VAC3φ50/60Hz220VAC3φ60Hz380VAC3φ50Hz400VAC3φ50Hz*6Utilityrequirements56A52A25A23A62dBAtambienttemperature+20℃,nospecimen.PerformanceshownaboveconformstoIEC60068-3-5:2001.Temperaturerateofchangeinthetemperaturerangeexcluding±10%ofmax/min.temperature.Noiselevel*7ExhaustheatquantitykJ/h(kcal/h)*1:*2:*3:Availableonrequest*4:*5:*6:ConformstoCEmarkingbasedonEUdirectives.*7:Measurementsaretobetakeninananechoicroomataheightof1.2mfromtheoorandadistanceof1mfromthechamberfrontpanelExcludingprotrusions.Powersupplyvoltageuctuationtobe±10%ofratedvalue.32400(7743)39600(9464)(ISO1996-1:2003_A-weightedsoundpressurelevel).


<< | < | > | >>