Platinous J Series Temperature & Humidity Chamber


>> P.24

INSTALLATIONREQUIREMENTSModelPRPHPPL12342341234PSL24PDR43PDL43PCR3PUPG123424200VAC3ø50/60Hz18.520.022.034.017.017.826.422.523.036.032.048.534.044.535.547.023.514.515.028.024.545.0Maximumcurrent(A)HumidierwatersupplyDrainageInstallationspace220VAC3ø60Hz*17.520.020.531.516.116.324.121.022.034.030.545.533.042.534.545.522.014.014.026.523.542.5380VAC3ø50Hz*8.58.010.019.59.519.08.68.315.410.011.022.018.030.017.527.018.529.011.09.010.513.517.523.0400VAC3ø50Hz*14.79.410.421.017.129.416.625.617.527.510.58.510.012.816.521.8Usepurewaterwithaconductivityof0.1to10μS/cmsuppliedfromthetank.─Drainportsarepositionedatthebottomoftherearpanel(150mmabovetheoor).Prepare1drainhosefortemperatureandhumidityuseand1drainhoseforcontinuouswatersupplyuse(option).Hoseouterdiameter:18mm,innerdiameter:12mmLength:approximately1mACBPR、PL、PUPSL、PGPHPPDR、PDLPCRType1Type2Type3Type4Type2Type4Type2Type3Type4Type3Type4Type3Spacetomanipulatethecableportandadjusterfeet,toconnectthepowersupplyandthewatersupplyanddrainpipes,andtoperformmaintenanceisrequired.(Werecommend30cmormore.)70801208012070801208012080Spacetopassthewaterdrainhosethroughandtoperformmaintenanceinisrequired.(Werecommend60cmormore.)60cmormoreAModelSide:AFront:B(cm)Rear:CTop*CompliancewithCEMarking*Thechamberdoesnotcomewithapowercable.A23


<< | < | > | >>