Trouble
|
Cause
|
Remedial action
|
|
AL00
CH00
|
The room temperature compensation
input to the temperature controller was disconnected.
The chamber has been stopped. |
The sensor is loosely connected
to the RTD terminal on the SCP-220(ANA-LOG) temperature
controller board or there is an open circuit
on the SCP-220 board. |
Turn control power OFF from the
POWER
key and resume testing.
If the same alarm occurs again, call for service. |
|
AL00
CH01 |
The dry-bulb input to the temperature
controller was disconnected. The chamber has been
stopped. |
The sensor is loosely connected
to the TD terminal on the SCP- 220(ANA-LOG)
temperature controller
boardor there is an open circuit in the connected
thermocouple. |
Turn control power OFF from the
POWER
key and resume testing.
If the same alarm occurs again, call for service. |
|
AL00
CH02 |
The wet-bulb/humidity input to
the temperature controller was disconnected. The
chamber has been stopped. |
The sensor is loosely connected
to the TW terminal on the SCP-220(ANA-LOG) temperature
controller boardor there is an open circuit
in the connected thermocouple. |
Turn control power OFF from the
POWER
key and resume testing.
If the same alarm occurs again, call for service. |
AL00
CH05
CH12 |
A refrigeration circuit monitor
channel (Ai5
12ch) to the temperature controller was disconnected.
The chamber has been stopped.
If the backup mode is ON
and the chamber has multiple refrigerators, only
the affected refrigerator stops while all others
continue running. |
Either the sensor is loosely connected
to the Ai5
12 terminal on the SCP-220(ANA-LOG) temperature
controller boardor there is an open circuit
in the connected thermocouple. |
Turn control power OFF from the
POWER
key (wait for the test end
if back up mode is ON) and resume testing. If
the same alarm occurs again, call for service. |
|
AL01 |
The temperature inside the chamber
has risen above the upper deviation limit. The
heater has been stopped until temperature returns
within range. |
Either specimens inside the chamber
are generating heat or the upper deviation limit
is set too low. |
Remove the heat-generating specimens
and/or set the upper deviation
limit about 10°C.
When temperature returns within range, normal
control is restored automatically. The alarm can
be cleared by pressing the Buzzer
key.
|
|
AL02 |
The temperature inside the chamber
has risen above the absolute high limit. The chamber
has been stopped. |
Either specimens inside the chamber
are generating heat or the absolute high limit
is set too low. |
Turn control power OFF from the
POWER key. Then, remove
the heat-generating specimens and/or set
the absolute high limit about 15°C
higher than the target temperature.
Resume testing. If the same alarm occurs again,
call for service.
|
|
AL03 |
The temperature inside the chamber
has dropped below the absolute low limit. The
chamber has been stopped. |
Possible causes include infiltrating
of outside air, excessive cooling, the effects
of cooling sources inside the chamber, and the
absolute low limit being set too high. |
Turn control power OFF from the
POWER
key and check the chamber
door and cable ports. If
the refrigerator is on manual control, correct
the capacity setting. If cooling sources
are overcooling the chamber, reduce their number.
Also, correct the absolute low limit setting as
necessary. Then, resume testing. If the same alarm
occurs again, call for service. |
|
AL06 |
The temperature inside the chamber
has risen above the overheat protector setting
(instrumentation panel). The chamber has been
stopped. |
Either specimens inside the chamber
are generating heat or the overheat protector
is set too low. |
Turn control power OFF from the
POWER
key. Then, remove the heat-generating
specimens and/or correct the overheat protector
setting. Resume testing. If the same alarm occurs
again, it is possible the thermal fuse has blown.
Call for service. |
|
AL07 |
The built-in temperature switch
of the air circulator motor (water circuit [heat
exhaust] compartment) tripped because the motor
is abnormally hot. The chamber has been stopped.
|
Overloaded air circulator motor |
Leave the chamber OFF until the
air circulator cools down. Then, turn control
power ON from the POWER
key and resume testing. |
AL08 |
The high pressure switch tripped
because refrigerator pressure was high. The chamber
has been stopped.. |
The refrigerator's operating pressure
is higher than specified. |
Turn control power OFF from the
POWER
key. With air-cooled refrigerators,
check if the condenser
filter is clogged. Correct as necessary and
resume testing. If the same alarm occurs again,
call for service. |
|
The thermal relay tripped because
refrigerator operating current was high. The chamber
has been stopped. |
The refrigerator's operating current
is higher than specified. |
Turn control power OFF from the
POWER
key. With air-cooled refrigerators,
check if the condenser
filteris clogged. Correct as necessary and
resume testing. If the same alarm occurs again,
call for service. |
|
The compressor temperature switch
tripped because refrigerator surface temperature
was high. The chamber has been stopped. |
The refrigerator's surface temperature
is higher than specified. |
Turn control power OFF from the
POWER
key. With air-cooled refrigerators,
check if the condenser
filteris clogged. Correct as necessary and
resume testing. If the same alarm occurs again,
call for service. |
|
The thermal relay of the condenser
fan tripped because fan operating current was
high. The chamber has been stopped. |
The condenser fan motor operating
current is higher than specified. |
Turn control power OFF from the
POWER key and check if
the condenser filter is clogged. Correct as necessary
and resume testing. If the same alarm occurs again,
call for service. |
AL08
CH05
CH12 |
One of the temperature controller
channels (Ai5
12ch) that monitor the refrigeration circuit was
disconnected. The chamber has been stopped. If
the chamber has more than 1 refrigerator and the
backup mode is ON, only the affected refrigerator
will shut down, while the chamber will continue
running using another refrigerator. |
Frost buildup, compressor trouble
or a refrigerant leak are all possible causes.
|
After testing ends, turn control
power OFF from the POWER
key. With air-cooled refrigerators,
check if the condenser
filter is clogged. Correct as necessary and
resume testing. If the same alarm occurs again,
call for service. |
|
AL11 |
The heater's circuit protector
tripped because operating current was high. The
chamber has been stopped. |
Short-circuit or overcurrent in
the heater circuit |
Set the main power switch (leakage
breaker) in the OFF position and
reset the humidifier's circuit
protector on the electric parts chassis(electric
parts compartment). Then, reactivate the system
from the main power switch followed by the POWER
key, and resume testing.
If the same alarm occurs again, call for service.
|
|
AL19 |
A reverse or open phase was detected
in the 3 primary
power supply connection. The chamber has been
stopped.
* A reverse or open phase
can be detected even when the POWER
key is in the OFF position. |
Primary power supply is not correctly
connected to the chamber. |
An incorrect power supply can
seriously affect chamber components.
Set the main power switch
(leakage breaker) in the OFF position, and check/correct
primary power supply phase alignment and connections.
Then, reactivate the system from the main power
switch followed by the POWER
key, and resume testing.
If the same alarm occurs again, call for service. |
|
Trouble was detected in an external
unit. The chamber has been stopped. |
For information on external unit
troubles, see the instruction manual that came
with the external unit. |
Turn control power OFF from the
POWER
key. Then, take the remedial
action described in the instruction manual of
the external unit. Resume testing. If the same
alarm occurs again, call for service. |
|
AL21 |
The humidifier's circuit protector
tripped because operating current was high. The
chamber has been stopped. If the backup mode is
ON, humidity tests are suspended, but temperature
tests continue. |
Short-circuit or overcurrent in
the humidifier circuit. |
Set the main power switch (leakage
breaker) in the OFF position and
reset the humidifier's circuit
protector on the electric parts chassis
(electric parts compartment). Then, reactivate
the system from the main power switch followed
by the POWER
key, and resume testing.
If the same alarm occurs again, call for service.
|
|
AL22 |
The humidity inside the chamber
has risen above the absolute high limit. The humidifier
has been stopped until humidity returns within
range. |
The absolute high limit is set
too low. |
Set
the absolute high limit setting about 10%RH higher
than the humidity SV.
When humidity returns within range, normal control
is restored automatically. The alarm can be cleared
by pressing the Buzzer
key. |
|
|
The humidity inside the chamber
has dropped below the absolute low limit. The
heater has been stopped and refrigerator capacity
decreased to the minimum (turned OFF in some cases)
until humidity returns within range. |
The absolute low limit is set
too high. |
Set
the absolute high limit setting about 10%RH higher
than the humidity SV.
When humidity returns within range, normal control
is restored automatically. The alarm can be cleared
by pressing the Buzzer
key. |
|
AL26 |
Humidifying tray water level does
not lower even though water is being drained.
The chamber has been stopped. |
Either the drainage pump has malfunctioned
or the drainage line is clogged. |
Check the water circuit drainage
pump (water circuit compartment) is functioning
correctly, and check for clogs in the drainage
line and clean as necessary. Then, resume testing.
If the same alarm occurs again, call for service.
|
|
The main water tank is empty.
The humidifier has been stopped. |
The main water tank is empty.
|
Refill the auxiliary water tank
and set it on the main water tank. Then, resume
testing. After water has been sufficiently supplied
to the main water tank, the alarm can be cleared
by pressing the Buzzer
key.
If the same alarm occurs
again, call for service. |
|
The humidifying tray did not fill
within the specified amount time at the start
of humidity testing. The chamber has been stopped.
If the backup mode is ON, humidity tests are suspended,
but temperature tests continue. |
The strainer on the humidifying
water supply circuit is either clogged or the
circuit is leaking, or water supply pressure is
too low. |
Check the humidifying water supply
circuit (wait for the test end if the backup mode
is on) and correct as necessary. Then, resume
testing. If the same alarm occurs again, call
for service. |
|
The humidifying tray did not fill
within the specified amount time during humidity
testing. The chamber has been stopped. If the
backup mode is ON, humidity tests are suspended,
but temperature tests continue. |
The strainer on the humidifying
water supply circuit is either clogged or the
circuit is leaking, or water supply pressure is
too low. |
Check the humidifying water supply
circuit (wait for the test end if the backup mode
is on) and correct as necessary. Then, resume
testing. If the same alarm occurs again, call
for service. |
|
During humidity testing, the wet-bulb
(measures relative humidity) rose above the specified
temperature. The humidifier will continue as is,
but humidity control might destabilize and trigger
a humidity alarm. |
The wet-bulb wick (gauze) inside
the chamber may be dry. |
Stop the chamber and replace the
wick. After wet-bulb temperature returns within
the specified range, the alarm can be cleared
by pressing the Buzzer
key. |
|
The auxiliary water tank is empty.
The humidifier will continue on the water in the
main tank, but it will eventually stop unless
water is replenished. |
The auxiliary water tank is empty.
|
Refill
the auxiliary water tank and resume testing. After
water has been sufficiently supplied to the main
water tank, the alarm can be cleared by pressing
the Buzzer
key.
If the same alarm occurs again,
call for service. |
|
AL99 |
The instrumentation detected communication
trouble between the CPU board and display board.
Or the instrumentation detected system trouble.
The chamber has been stopped. |
Communication error between the
CPU board and display board |
Reactivate the system from the
main power switch (leakage breaker) and resume
testing. If the same alarm occurs again, call
for service. |