Platinous K series(T-instrumentation):When an Alarm Occurs |
| |
| The following procedure explains what to do when one of the chamber's self-check features trips. |
 |
|
|
Procedure |
 |
1. |
Press the BUZZER key on the instrumentation panel to silence the buzzer. |
|
 |
| 2. |
If the chamber is still running, press the OPER./STOP key to stop testing. |
 |
| 3. |
If control power is still ON, press the POWER key on the instrumentation panel to shut it OFF. |
 |
| 4. |
Set the main power switch in the OFF position. |
 |
| 5. |
Check the displayed alarm code and remedy the trouble as explained in the following alarm table. |
 |
| 6. |
To resume testing, first set the main power switch in the ON position then press the POWER key. |
|
 |