Platinous K series(P-instrumentation):When an Alarm Occurs

The chamber will be in one of the below states when an alarm occurs. The alarm is cleared in a different way depending on what your needs are at that time. Procedures for clearing the alarm in each of these cases are given on the following pages.
• Running in the constant mode
• Running in the program mode but not needing to continue after clearing the alarm
• Running in the program mode and needing to continue after clearing the alarm
When running in the constant mode or when running in the program mode but not needing to continue after clearing the alarm
Procedure 1. Press the Buzzer Stop button on the Alarm screen to silence the buzzer.
2. If the chamber is running, press either the OPER./START key on the operating panel or the chamber operating status box on the screen to get the Operation Mode Selection screen. Then, press the STOP button under Stop Operation.
3. Press the POWERkey to deactivate control power.
4. Set the main power switch in the OFF position.
5. Remedy the trouble for the alarm displayed on the Alarm screen as explained in the alarm table.
6. Set the main power switch in the ON position and then activate control power. The display will come on and the main menu will appear shortly.
When running in the program mode and needing to continue after clearing the alarm
Procedure 1. Press the Buzzer Stop button on the Alarm screen to silence the buzzer.
2. Press thePOWER key to deactivate control power.Check the alarm table. If there is no need to shut OFF primary power supply, leave the main power switch in the ON position.
3. Remedy the trouble for the alarm displayed on the Alarm screen as explained in the alarm table.
4. Reactivate control power.
5. If main power switch was left in the ON position, a message confirming whether to resume testing from the state prior to the alarm or not will appear on the screen. If the main power switch was set in the OFF position, set it in the ON position and then activate control power to start the chamber.
Yes  :  Resumes testing from the state prior to the alarm.
No :  Leaves the chamber stopped.