Product FAQ/Day-to-day maintenance and repair methods

Back to Products FAQ List

 

Compact Ultra Low Temperature Chamber
(T controller):FAQ

Procedure 1. Press either the or keys to silence the buzzer.
2. Check the alarm code?iAL XX?jon the display.
3. If the chamber is running, press the OPER./STOP key to stop the chamber.
4. If control power is ON, press the POWER key to shut OFF power to the instrumentation.
5. Set the main power switch in the OFF position.
6. Remedy the trouble as explained in the Alarm Tables.
7. To resume operation, set the main power switch in the ON position, press the POWER key and then press the OPER./STOP key in that order.