Problem
Cause
Action
AL00
Controller CH?PBurn-out
Dry bulb sensor is disconnected. • Loose connection on the terminal 25 or 26 in ?d?r-101-CPU.
• The dry bulb sensor is disconnected.
• Turn off the power switch.
• Check the input terminal ES-101-CPU
AL01
High-limit (deviation)Warning
Actual temperature of the inside of the chamber is exceeding the upper limit deviation setting value. • There is too much live load.
• The setting of the upper deviation is too low.
• Remove the live load
.• Set the setting value 10°C higher than previous setting.
• This error is reset when actual temperature get back the limit.
AL02
High-limit error
Actual temperature of the inside of the chamber is exceeding the upper limit absolute setting value • There is too much live load.
• The setting value of lower limit absolute setting is too low
• Turn off the power switch.
• Remove the live load.
• Check the high-limit setting value.
AL03
Low-limit error
Actual temperature of the inside of the chamber is exceeding the lower limit. • Too much cooling power due to wrong setting in the Ref.
• test sample pull down the temperature.
• The setting of the lower limit is too high.
• Turn off the power switch.
• Remove the test sample.
• Check the setting of the lower limit.
• This error is reset when actual humidity get back the limit.
AL06
Over-Heat error
Actual temperature of the inside of the chamber is exceeding the setting value in the over heat protector. • There is too much live load.
• The setting value of the over heat protector is too low.
• Turn off the power switch.
• Remove live load.
• Check the setting in the over heat protector.
AL07
Air circulator error
The current of the air circulation motor is exceeding the limit and over load relay is activated. • Over load operation for the air circulation motor. • Turn off the power switch.
• Stop the operation and cooling down the motor.
AL08
Ref. Error
Cooling water pressure become less than setting of water pressure switch.
[Only water cooling type.]
All the refrigeration units have a problem.
• The pump for cooling water line does not work.
• The strainer filter has clogging.
• water leaking in the cooling water line.
• Each single alarm message has been shown.
• Turn off the power switch
• Check the cooling water pump.
• Check the water leakage in cooling water line.
Cleaning up the strainer filter.
AL11
Heater error
The over current protector is activated due to over current • The heater makes short circuit or over current. Turn off the main breaker.
Reset the breaker for the heater to reset the error
AL12
Air Con. Local overheating
The temperature switch of the air conditioning unit is activated due to high temperature. • There is too much live load.
• Malfunction of the heater.
• Malfunction of the motor.
• Turn off the main breaker after High temperature operation.
• Turn off the power switch.
• Blow the fresh air into the air conditioning unit to pull down the temperature below 50°C
AL13?`16
No.1?`4
Ref. error
• Temperature switch of the compressor is activated
• High pressure switch is activated.
• Over load relay of the compressor is activated.
• Malfunction of the compressor.
• Malfunction of the condenser.
• Leakage of the refrigerant.
• Super heat
• Abnormal phase.
• Turn off the power switch.
• Cooling down the compressor.
Cleaning up the strainer filter.
AL20
Controller CH2 Burn-out
Wet bulb input has miss-connection on the controller or wet bulb sensor is disconnected. • The terminal No. 31, 32, 33 and 34 at ES-101-CPU is disconnected.
• Disconnection of the sensor.
• Turn off the power switch.
AL21
Humidifier error
Over current protector for humidifier is activated due to over current • Humidifier electric circuit makes short circuit or over current. Turn off the main breaker.
Reset the breaker for the humidifier.
AL22
High-limit absolute error
Actual humidity of the inside of the chamber is exceeding the upper limit absolute setting value • The setting value of upper limit absolute setting is too low. • Check the setting value of upper limit absolute setting
• This error is reset when actual humidity get back the limit.
AL27
Humidifier Boil-Dry
Inside temperature of the humidifier rise up and exceeding and activate the temperature switch. • Malfunction of the water circuit of the humidifier.
• There is many scale in the humidifier.

• Turn off the power switch.
Clean up the humidifier.

AL28
Humidifier water supply Error.
Humidifier can not supply the water 100% even after get the “humidifier water supply Warning • Safety device for DI water supply system such as RO water supply is activated. • Turn off the power switch.
• Check the water valve.
Clean up the strainer filter.
Check the water supply line.
• Check the safety device of DI water supply unit.
AL29
Humidifier Drain error
Humidifier cannot drain the humidifier water. • Malfunction of the drain water pump.
• Clogging of the drain line.
• Malfunction of the float switch.
Clean up the humidifier
AL30
Humidifier Water level float error
Malfunction of the float switch in the humidifier • Malfunction of the float switch • Turn off the power switch.
AL42
Output circuit error.
The over current protector is activated due to over current • The controlling circuit makes short circuit or over current Turn off the main breaker.
Reset the breaker for the humidifier to reset error
AL45
Sequencer error
Malfunction of the sub-unit of the controller.
(This error occurs even the chamber is not running.)
• Sub-unit has CPU FAIL
• Disconnection of the cable on the sub unit.
Turn off the main breaker.
AL50
Scale in Humid.: AccumulationWarning.
Temperature switch is activated due to over heat. • There is many scale in the humidifier. Clean up the humidifier.
AL51
Humidifier water supply Warning.
Humidifier can not supply the water within limit • Water supply valve is closed.
• Strainer filter is clogged.
• Water leaking
• Water supply pressure is reducing.
• Malfunction of the float switch for humidifier.
• Check the valve.
Clean up the strainer filter.
Check the water supply line.
This problem is automatically reset after humidifier got enough water.
AL52
Check humidifier operation.
Humidifier ON-OFF switch set to “CLEAN” during humidity operation. Forget to set the Humidifier ON-OFF switch for “ON”When set this switch to “ON” or press clear key in the display, humidifier starts to supply the water Set the Humidifier ON-OFF switch to “ON” after cleaning