ESPEC


Home : Products Information : Evaluation Systems
Select by market
Evaluation Systems
TDDB Evaluation System
Catalog Download
PDF509KB

About Specification
MSM Simplex Performance
 
Voltage range Resolution Accuracy Max current
±10V 1mV ±(0.2%+10mV) 100mA
±50V 10mV ±(0.2%+50mV) 100mA
 
Current range Resolution Accuracy Max current
±100mA 100μA ±(0.5%+100μA+2μAx Vo) 50V
±10mA 10μA
±(0.5%+10μA+200nAx Vo)
±1mA 1μA ±(0.5%+1μA+20nAx Vo)
±100μA 100nA
±(0.5%+100nA+2nAx Vo)
±10μA 10nA ±(1.0%+10nA+200pAx Vo)
±1μA 1nA
±(1.0%+1nA+20pAx Vo)
±100nA 100pA ±(1.0%+100pA+2pAx Vo)
±10nA 10pA ±(2.0%+10pA+200fAx Vo)
±1nA 1pA ±(2.0%+1pA+20fAx Vo)
Accuracy: ±(set value or % of specified value) ±(offset), Vo: output voltage (V)
 
 
Model AMM-1000
Application OS: Windows NT TDDB Evaluation System
Voltage/current
application range
-50V to +50V/-100mA to +100mA
Resolution 1mV step/1pA step
Voltage/current
measurement range
-50V to +50V/-100mA to +100mA
Number of
measurement channels
Standard 40ch. Max. installment 200ch
Measurement
sampling speed
By measurement averaging setting
Short mode: 0 to 100msec →10msec interval
100msec to 100sec →100msec interval
Over 100sec → 1sec interval

Medium mode: 0 to 100msec →20msec interval
100msec to 100sec →100msec interval
Over 100sec → 1sec interval

Long mode: 100msec interval
Short mode: Data acquisition without averaging
Medium Long mode: Averaging per 1cycle
External dimension
of system
System controller W650x H1300x D800mm
Control rack W570x H1100x D900mm
Required utility AC100V±10% 50/60Hz 15Ax l
AC100V±10% 50/60Hz 50Ax l

Copyright (C) ESPEC CORP.All rights reserved