Capasitor / Inductor Temperature Characteristic Evaluation System

Overview of Equipment

The Capacitor Temperature Characteristic Evaluation System automates temperature step control of the standard environmental test system and measurement of multiple channels of condenser electrostatic capacity, dielectric loss tangent (tanδ), and impedance. This enables automatic recording of data of frequency characteristics and change over time under desired temperature environments. Not only does it evaluate the characteristics of the condenser, but it can also evaluate various electronic parts, print substrates, and insulation material, and it can be used in combination with your own standard environmental test system.

Automatic measurement of up to 64 channels
This system can measure the multiple channels of electrostatic capacity, dielectric loss tangent (tanδ), and impedance under a temperature and humidity environment. There are 8 standard channels, and it can be expanded to a maximum 64 channels.
Choose from 3 test modes
Choose from 3 test modes: temperature (and humidity) characteristic testing, frequency characteristic testing, and life characteristic (constant operation) testing to evaluate changes over time.
Features
  • Temperature characteristic evaluation testing: Automatically records characteristic data synchronized with changes in temperature up to 20 steps.
  • Frequency characteristic evaluation testing: Automatically records characteristic data for various frequencies while changing the frequency in a temperature and humidity environment.
  • Life characteristic (constant operation) testing: Automatically measures and records changes over time of characteristics in a temperature and humidity environment.

Please scroll horizontally to look at table below.

Measurement method Alternating current 4-terminal measurement (measuring cable tip)
Measurement range Measurement frequency: 20 Hz to 1 MHz
Dielectric loss tangent: 0.0001 to 10.0000 tanδ
Electrostatic capacity: Depends on impedance measurement range
Impedance: 10 mΩ to 100 MΩ
Measurement range Select from AUTO, 10Ω, 100Ω, 300Ω,
1kΩ, 3kΩ, 10kΩ, 30kΩ, or 100kΩ
Measurement interval Min. 1 min. to 1500 min. (Variable in 1 min. increments)

Detailed Specifications

Detailed specifications

Please scroll horizontally to look at table below.

Model AMQ-008-C
Measurement items Electrostatic capacity, dielectric loss tangent (tanδ), and impedance (Select 2 for the data to be recorded)
Test mode
  • Temperature characteristic evaluation testing (change with respect to temperature)
  • Frequency characteristic evaluation testing (change with respect to frequency)
  • Humidity characteristic evaluation testing (change with respect to humidity with constant temperature)
  • Constant operation testing (change with respect to test time)
  • Temperature and humidity characteristic evaluation testing (change with respect to temperature and humidity)
Channel configuration Standard 8 channels, maximum 64 channels expandable in 8 channel increments
Measurement control Measurement method: Alternating current 4-terminal measurement (measuring cable tip)
Measurement range:
  • Measurement frequency: 20 Hz to 1 MHz
  • Dielectric loss tangent (tanδ): 0.0001 to 10.0000
  • Electrostatic capacity: Depends on impedance measurement range
  • Impedance: 10 mΩ to 100 MΩ
Measurement instrument LCR meter (with E4980A 4m compensation option)
Measurement range Select from AUTO, 1Ω, 10Ω, 100Ω, 300Ω, 1kΩ, 3kΩ, 10kΩ, 30kΩ, or 100kΩ
DC bias 0 V, 1.5 V, 2.0 V (optional max. 40 V)
Measurement interval: Min. 1 min. to 1500 min. (Variable in 1 min. increments)
Temperature steps Select from mode that specifies 20-step start/end temperature and step interval and mode for entering desired temperature
Frequency steps 20 steps (desired value can be set)
Compensation Short compensation, open compensation
Standard environmental test system control
  • Enables control of temperature and humidity of standard environmental test system with RS-485 function and loading of temperature and humidity data synced with measurement.
Measuring cable Teflon coaxial cable 4 m (characteristic impedance 50 Ω, 95 PF/m)
Utility Requirements 100 V AC power supply, single phase 15 A
Options
  • Relay unit
  • Jig (chip type, lead type, direct mount type)
  • LAN supported software
  • Insulation resistance measurement function
  • Sampling evaluation software