Platinous J Series Temperature & Humidity Chamber


>> P.34

OptionsPerformanceSpecimentemperaturecontrolTemp.&humid.SPattainmentoutputSensorsareattachedtothespecimentoallowexposureteststhatprovideaccuratetemperaturestresstothespecimen.•Insulatedtype•Non-insulatedtypeWhenthetemperature(humidity)inthechamberreachesthesetvalues,thechambersendsoutacontactsignal.Itsynchronizesthepowersupplytothespecimen,thetimingformeasurementsortopreventdewfromcondensingonthespecimens.ApplyingDCpowersupplyCapableofapplyingvoltagetothespecimen,usedforbiastesting.TheDCpowersupplyunitsynchronizeswithconstantandprogramoperations,andcanbesetforeachtemperatureandhumidityprogramstep.RatedvoltageRatedcurrentVoltagesettingrange5V60A1.0to5.5V12V27A2.4to13.2V15V22A3.0to16.5V24V14A4.8to26.4V48V7A9.6to52.8VCapacitivehumiditysensorAttachedinplaceofthewetbulbwick.Measurementrange:0%rhto100%rhAccuracy:±2%rh(-20℃to+40℃and0%rhto90%rh)TimeupoutputThisoptionenablesturningthepowertothespecimenONorOFFwithcontactsignaloutputwhenthetimeisupbyusingthetimerfunctiononthetemperature(humidity)controller.TimesignalterminalAddsadditionalterminalstothestandardtimesignalterminals.33


<< | < | > | >>